ADUC7036DCPZ-RL Analog Devices Inc, ADUC7036DCPZ-RL Datasheet - Page 53

no-image

ADUC7036DCPZ-RL

Manufacturer Part Number
ADUC7036DCPZ-RL
Description
SFlash 96k ARM7 Dual 16-Bit ADC LIN I.C.
Manufacturer
Analog Devices Inc
Series
MicroConverter® ADuC7xxxr
Datasheet

Specifications of ADUC7036DCPZ-RL

Core Processor
ARM7
Core Size
16/32-Bit
Speed
20.48MHz
Connectivity
LIN, SPI, UART/USART
Peripherals
PSM, Temp Sensor, WDT
Number Of I /o
9
Program Memory Size
96KB (96K x 8)
Program Memory Type
FLASH
Eeprom Size
-
Ram Size
6K x 8
Voltage - Supply (vcc/vdd)
3.5 V ~ 18 V
Data Converters
A/D 2x16b
Oscillator Type
Internal
Operating Temperature
-40°C ~ 115°C
Package / Case
48-VFQFN Exposed Pad, CSP
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Current Channel ADC Data Register
Name: ADC0DAT
Address: 0xFFFF0520
Default Value: 0x0000
Access: Read only
Function: This ADC data MMR holds the 16-bit conversion
result from the I-ADC. The ADC does not update this MMR if
the ADC0 conversion result ready bit (ADCSTA[0]) is set. A
read of this MMR by the MCU clears all asserted ready flags
(ADCSTA[2:0]).
Voltage Channel ADC Data Register
Name: ADC1DAT
Address: 0xFFFF0524
Default Value: 0x0000
Access: Read only
Function: This ADC data MMR holds the 16-bit voltage
conversion result from the V-/T-ADC. The ADC does not
update this MMR if the voltage conversion result ready bit
(ADCSTA[1]) is set. If I-ADC is not active, a read of this MMR
by the MCU clears all asserted ready flags (ADCSTA[2:1]).
Temperature Channel ADC Data Register
Name: ADC2DAT
Address: 0xFFFF0528
Default Value: 0x0000
Access: Read only
Function: This ADC data MMR holds the 16-bit temperature
conversion result from the V-/T-ADC. The ADC does not
update this MMR if the temperature conversion result ready bit
(ADCSTA[2]) is set. If I-ADC and V-ADC are not active, a read
of this MMR by the MCU clears all asserted ready flags
(ADCSTA[2]). A read of this MMR clears ADCSTA[2].
Current Channel ADC Offset Calibration Register
Name: ADC0OF
Address: 0xFFFF0530
Default Value: Part specific, factory programmed
Access: Read/write
Function: This ADC offset MMR holds a 16-bit offset calibration
coefficient for the I-ADC. The register is configured at power-
on with a factory default value. However, this register automati-
cally overwrites if an offset calibration of the I-ADC is initiated
by the user via bits in the ADCMDE MMR. User code can write
to this calibration register only if the ADC is in idle mode. An
ADC must be enabled and in idle mode before being written to
any offset or gain register. The ADC must be in idle mode for at
least 23 μs.
Rev. C | Page 53 of 132
Voltage Channel ADC Offset Calibration Register
Name: ADC1OF
Address: 0xFFFF0534
Default Value: Part specific, factory programmed
Access: Read/write
Function: This offset MMR holds a 16-bit offset calibration
coefficient for the voltage channel. The register is configured at
power-on with a factory default value. However, this register is
automatically overwritten if an offset calibration of the voltage
channel is initiated by the user via bits in the ADCMDE MMR.
User code can write to this calibration register only if the ADC
is in idle mode. An ADC must be enabled and in idle mode
before being written to any offset or gain register. The ADC
must be in idle mode for at least 23 μs.
Temperature Channel ADC Offset Calibration Register
Name: ADC2OF
Address: 0xFFFF0538
Default Value: Part specific, factory programmed
Access: Read/write
Function: This ADC offset MMR holds a 16-bit offset
calibration coefficient for the temperature channel. The register
is configured at power-on with a factory default value. However,
this register is automatically overwritten if an offset calibration
of the temperature channel is initiated by the user via bits in the
ADCMDE MMR. User code can write to this calibration register
only if the ADC is in idle mode. An ADC must be enabled and
in idle mode before being written to any offset or gain register.
The ADC must be in idle mode for at least 23 μs.
Current Channel ADC Gain Calibration Register
Name: ADC0GN
Address: 0xFFFF053C
Default Value: Part specific, factory programmed
Access: Read/write
Function: This gain MMR holds a 16-bit gain calibration
coefficient for scaling the I-ADC conversion result. The register
is configured at power-on with a factory default value. However,
this register is automatically overwritten if a gain calibration of
the I-ADC is initiated by the user via bits in the ADCMDE
MMR. User code can write to this calibration register only if the
ADC is in idle mode. An ADC must be enabled and in idle
mode before being written to any offset or gain register. The
ADC must be in idle mode for at least 23 μs.