8845A/SU 120V Fluke Electronics, 8845A/SU 120V Datasheet - Page 38

PREC MLTMTR DMM BNCH 6.5 DIG RES

8845A/SU 120V

Manufacturer Part Number
8845A/SU 120V
Description
PREC MLTMTR DMM BNCH 6.5 DIG RES
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8845A/SU 120V

Includes
Test Leads
Style
Bench
Display Digits
6.5
Display Type
VFD, Dual
Display Count
Varies
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Memory, RS-232
Ranging
Manual
Response
Average
Lead Free Status / RoHS Status
Contains lead / RoHS compliant by exemption
Other names
2675292
614-1001
8845A/8846A
Users Manual
2-6
To replace the Current Input fuses,
1. Turn the Meter off, unplug the power cord from the Meter, and remove all test
2. Turn the Meter on its back.
3. Unscrew the retaining screw on the fuse access door as depicted in Figure 2-2.
4. Remove the protective cover from the fuse holders by slightly depressing the
5. Remove the defective fuse, and replace with one having the appropriate rating.
6. Replace the protective cover by pushing it over the fuses while aligning the
7. Replace the fuse door and secure it by tightening the retaining screw.
If the fuse is good, the Meter will read less than 1 . If the fuse is blown, the
Meter will read overload.
leads.
back edge of the cover to unlatch it from the printed circuit board. Pull up on the
back edge of the cover and remove it from the fuse compartment.
catches with the holes in the printed circuit board. Press the cover down until the
catches engage the printed circuit board.
Figure 2-2. Replacing the Current Input Fuses
Fuses
F2
F1
Bottom front
left corner
caw020.eps