1141A AGILENT TECHNOLOGIES, 1141A Datasheet - Page 10

Differential Oscilloscope Probe

1141A

Manufacturer Part Number
1141A
Description
Differential Oscilloscope Probe
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 1141A

Bandwidth
200MHz
For Use With
Agilent Technologies 6000 & DSO5000 Series Oscilloscopes
Lead Free Status / RoHS Status
na
High-voltage Differential Active Probes (continued)
N2792A 200-MHz and N2793A 800-MHz
general-purpose differential probe
The N2792A 200-MHz and N2793A 800-MHz differential
probes provide the superior general-purpose differential
signal measurements required for today’s high-speed
power measurements, vehicle bus measurements, and
digital system designs.
The N2792A and N2793A probes offer a 10:1 attenuation
setting and high input resistance and low input
capacitance to minimize circuit loading.
Both probes are compatible with any oscilloscope with
50 Ω BNC input. The probe can be powered by any USB
port on a scope or computer, or by an internal battery.
Characteristics for N2792A and N2793A
differential probes
Ordering information for Agilent differential probes and
power supply
Bandwidth
Risetime
Attenuation ratio
CMRR
Max input voltage
to ground
Max input voltage
between two
inputs
Max number of
probes supported
by 3000 X/5000/
6000/7000 Series
1141A
1142A
N2790A
N2791A
N2792A
N2793A
N2891A
200-MHz differential probe
Probe control and power module for 1141A
100-MHz, 1.4 kV differential probe with AutoProbe
interface
N2791A 25-MHz, 700-V differential probe
N2792A 200-MHz, 20-V differential probe
N2793A 800-MHz, 15-V differential probe
N2891A 70-MHz, 7,000-V differential probe
N2792A
200 MHz
1.75 ns
10:1
80 dB at 50/60 Hz
-50 dB at 10 MHz
±60 V
±20 V
4
N2793A
800 MHz
437 ps
10:1
-60 dB at 50/60 Hz
-15 dB at 500 MHz
±40 V
±15 V
4
10
N2890A 100-MHz, 1.4-kV differential probe with
AutoProbe interface
N2791A 25-MHz, 700-V differential probe
N2792A 200-MHz, 20-V differential probe
N2793A 800-MHz, 15-V differential probe