1141A AGILENT TECHNOLOGIES, 1141A Datasheet - Page 12

Differential Oscilloscope Probe

1141A

Manufacturer Part Number
1141A
Description
Differential Oscilloscope Probe
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 1141A

Bandwidth
200MHz
For Use With
Agilent Technologies 6000 & DSO5000 Series Oscilloscopes
Lead Free Status / RoHS Status
na
Single-ended Active Probes (continued)
Agilent 1156A single-ended active voltage
• 1.5 GHz probe bandwidth;
• 100 kΩ, 0.8 pF non-resonant input
• Small size for easier probing
• Compatible with 5000, 6000, and
The Agilent 1156A is a 1.5-GHz
active probe for use with
Agilent InfiniiVision or Infiniium
oscilloscopes. Its high bandwidth
(1.5 GHz), low input capacitance
(< -0.8 pF) and high input resistance
(100 kΩ) input minimizes ultra low
loading of the DUT, making it ideal for
the companion for the InfiniiVision
Series oscilloscopes. When used with
the InfiniiVision 3000 X, 5000, 6000,
or 7000 Series, the 1156A offers you
a full bandwidth of the scope to the
probe tip, giving you the accurate
insight into your high-speed device.
Incorporating the most advanced
mechanical and electrical design
technologies, the 1156A provides
smaller, lighter, and more rugged
probe tip that enables direct and
easy access to fine pitch ICs and
components. With the 1156A probe,
a damping resistor is placed as close
as possible to the point being probed,
which keeps the input impedance
from resonating low, and it also
allows a flat frequency resonance
across the entire bandwidth of the
probe. The 1156A is designed to
withstand 40-V peak AC input and
> -5 kV of ESD, so it will function
reliably in adverse condition.
< 233 ps rise/fall time
impedance
7000 Series (except for 6000 Series
100 MHz models)
Note
Ordering information for Agilent 1156A active probe
Agilent 1156A active probe with
accessories
Key characteristics
Probe bandwidth (-3 dB)
Rise and fall time (10 to 90%)
calculated from tr = 0.35/bandwidth
System bandwidth (-3 dB)
Input capacitance
Input resistance
Input dynamic range
DC attenuation
Offset range
Offset accuracy
Maximum input voltage
ESD tolerance
Maximum number of probes
supported by 3000 X/5000/6000/
7000 Series
1156A
Accessories
E2637A
E2638A
E2639A
E2640A
E2641A
1
: for waveforms with edges > 3 ns, dynamic range is > 12 Vp-p.
1.5 GHz single-ended active probe
Precision measurement kit (includes 2 solderable ground sockets with
2 green resistive signal pins)
Solderable-tip 5 cm resistive signal leads (10) with ground leads (3)
Micro clips, qty. 4
Resistive signal tips, qty. 8
Ground blade assembly, qty. 8
12
1
> 1.5-GHz
233 ps
1156A 1 GHz Infi niiVision 6000/7000
scope: 1 GHz
0.8 pF
100 kΩ ± 1%
5 Vp-p
10:1 ± 3% before calibration
10:1 ± 1% after calibration
±15.0 V
< 3% of setting before calibration
1% of setting after calibration
40 Vpeak, CAT I
> 5 kV from 100 pF, 300 Ω HBM
4
For more comprehensive information
about the 1156A active probe, refer
to the Agilent 1156/57/58A Active
Probes Product Overview, literature
number 5988-3361EN.