1141A AGILENT TECHNOLOGIES, 1141A Datasheet - Page 13

Differential Oscilloscope Probe

1141A

Manufacturer Part Number
1141A
Description
Differential Oscilloscope Probe
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 1141A

Bandwidth
200MHz
For Use With
Agilent Technologies 6000 & DSO5000 Series Oscilloscopes
Lead Free Status / RoHS Status
na
Single-ended Active Probes (continued)
• Up to 800-MHz bandwidth
• 10:1 attenuation
• Compatible with 3000 X, 5000, 6000
Agilent 1144A active probe
The 1144A features 800-MHz
bandwidth, 1-MΩ input resistance,
2-pF input capacitance, 10:1
attenuation, and ±40 Vdc + peak ac
maximum input voltage. An FET at the
input allows a high input resistance
and low input capacitance which
minimizes the loading of the circuit
under test. The output impedance
of the probe is 50 Ω which allows
the probe cable to be extended with
a 50-Ω coaxial cable. To use this
probe with 5000, 6000, or 7000 Series
oscilloscope, the 1142A power supply
is required. The 01144-61604 adapter
can be used with the power supply
to provide power for two channels of
active probing.
*
300 MHz - 1 GH, and 7000 Series
oscilloscopes
Operating characteristics
Characteristics for 1144A active probe
Bandwidth
Risetime
(calculated)
Attenuation ratio
Input resistance
Maximum input
voltage
Input
capacitance*
Input dynamic
range*
Power
requirements
Weight
Temperature
Humidity
Altitude
Maximum
number of probes
supported by
3000 X/5000/
6000/7000 Series
Net
Probe body
Shipping
Operating
Non-operating
Operating
Non-operating
Operating
Non-operating
13
≥ 800 MHz
≤ 440 ps
10:1, ±2% (when
connected to an
instrument input of
50 Ω, ±0.5%)
1 MΩ, ±5%
±40 V (dc + peak ac)
2 pF (typical)
0 to ±7.0 V
dc ±12 V to
±15 V ±5%
(at approximately
75 mA each supply)
Approximately
180 g (6.5 oz)
35 g (~1.25 oz)
Approximately
0.8 kg (1.75 lb)
0 to 55 °C
(32 to 131 °F)
-40 to +70 °C
(-40 to +158 °F)
Up to 95% relative
humidity (non-
condensing) at
40 °C (104 °F)
Up to 90% relative
humidity at
65 °C (149 °F)
Up to 4,600 m
(15,000 ft)
Up to 15,300 m
(50,000 ft)
4
1144A active probe