Current Probes (continued)
Characteristics of the 1146A
current probe
Note: Reference conditions 23 ± 5 °C,
(73.4 ± 41 °C) 20 to 75% relative humidity,
dc to 1 kHz, probe zeroed, 1-minute
warmup, batteries at 9 V + 0.1 V,
external magnetic fi eld <40 A/m, no dc
component, no external current carrying
conductor, 1 MΩ/100 pF load, conductor
centered in jaw.
*
Bandwidth*
Current
range*
Output signal 1000 mV peak max
AC current
accuracy*
Noise
Insertion
impedance
Maximum
working
voltage
Maximum
common
mode voltage
Infl uence
of adjacent
conductor
Infl uence of
conductor
position
Battery
Low battery
Battery life
Range
Accuracy
Range
Accuracy
Range
Accuracy
Characteristics marked with asterisks are
specifi ed performance. Others are typical
characteristics.
dc to 100 kHz (-3 dB)
100 mV/A:100 mA to
10 A peak
10 mV/A:1 to 100 A peak
100 mV/A
(50 mA to 10 A peak)
3% of reading ±50 mA
10 mV/A
(500 mA to 40 A peak)
4% of reading ±50 mA
10 mV/A
(40 A to 100 A peak)
15% max at 100 A
Range 10 mV/A: 480 µV
Range 100 mV/A: 3 mV
(50/60 Hz) 0.01 Ω
600 Vrms max.
600 Vrms max.
< 0.2 mA/A AC
0.5% of reading at 1 kHz
in jaw
9 V alkaline (NEDA
1604A, IEC 6LR61)
Green LED when > 6.5 V
55 hours typical
Characteristics of the 1147A/N2893A
current probe
Note: The above specifi cations are
guaranteed at 23 ± 3 °C (or 73 ± 5 °F)
Bandwidth
(–3 dB)
Risetime
Maximum
current
(continuous)
Maximum
peak current
(non-
continuous)
Output
voltage rate
Amplitude
accuracy
Noise
Temperature
coeffi cient
for sensitivity
Effect of
external
magnetic
fi elds
Maximum
rated power
Maximum
input voltage
Diameter of
measurable
conductors
Probe
interface
Cable
lengths
Maximum
number
of probes
supported
dc to 50 MHz (1147A)
dc to 100 MHz (N2893A)
7 ns or less
15 A peak, 15 A DC,
10 Arms
30 A peak
0.1 V/A
±1% rdg, ±10 mA (dc
and 45 to 66 Hz, rated
current)
Equivalent to 2.5 mArms
or less (for 20 MHz
bandwidth measuring
instrument)
±2% or less (within a
range of 0 to 40 °C or
32 to 104 °F)
Equivalent to a
maximum of 20 mA (in
a dc to 60 Hz, 400 A/m
magnetic fi eld)
3 VA (with rated
current)
300 V CAT I
5 mm dia. (0.2 in dia.)
AutoProbe interface
Sensor cable:
Appox. 1.5 m (59.0n in)
Power supply cable:
Appox. 1 m (39.4 in)
2
18
Characteristics of N2780A Series
current probes
Ordering information for Agilent
current probes
For more information about the
N2780A Series current probes, refer
to the Agilent N2780A Series current
probe data sheet, literature number
5989-6432EN.
Bandwidth
(-3 dB)
Maximum
current
(continuous)
Maximum
peak current
(non-
continuous)
Maximum
input voltage
Output
voltage rate
Amplitude
accuracy
1146A
1147A
N2893A 100-MHz current probe with
N2780A 2-MHz current probe
N2781A 10-MHz current probe
N2782A 50-MHz current probe
N2783A 100-MHz current probe
N2779A 3-channel power supply for
100-kHz current probe
50-MHz current probe with
AutoProbe interface
AutoProbe interface
N2780A/81A/82A/83A
dc to 2 MHz (N2780A)
dc to 10 MHz (N2781A)
dc to 50 MHz (N2782A)
dc to 100 MHz (N2783A)
500 A (N2780A)
150 A (N2781A)
30 A (N2782A/N2783A)
700 A peak (N2780A)
300 A peak (N2781A)
50 A peak
300 V CAT I (N2782A, 83A)
300 V CAT II, 600 V CAT III
(N2780A, 81A)
0.01 V/A
0.1 V/A
±1.0 % rdg ±500 mA
±1.0 % rdg ±100 mA
±1.0 % rdg ±10 mA
±1.0 % rdg ±10 mA
(N2783A)
(N2782A/N2783A)
(N2780A/N2781A)
(N2782A/N2783A)
(N2780A)
(N2781A)
(N2782A)