OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
Cleaning the Loss/Length Connectors
Always clean and inspect fiber connectors before making
connections.
Cleaning the Loss/Length OUTPUT Connector
1
2
3
4
5
7-4
Touch the tip of a fiber optic solvent pen or swab
soaked in solvent to a lint-free dry wipe or fiber
cleaning card.
Touch a new, dry swab to the solvent spot on the wipe
or card.
Push the swab into the connector, twist it around 3 to 5
times against the end-face, then remove and dispose of
the swab.
Dry the connector with a dry swab by twisting it
around in the connector 3 to 5 times.
Inspect the connector regularly with a fiber
microscope, such as the Fluke Networks FiberInspector
Video Microscope.
Cleaning the Loss/Length and Power Meter INPUT
Connector
1
2
Verifying Operation
When you install the loss/length module in the tester, the
module’s model number should show in the upper-right
corner of the screen. If the screen shows No Module
Installed, refer to “If Something Seems Wrong” in Chapter
10.
Loss/Length Test Settings
Table 7-1 summarizes the loss/length test settings. To access
these settings, press
Remove the connector adapter to expose the
photodiode lens (see Figure 7-2).
Use the method described above for the loss/length
OUTPUT connector to clean the lens.