34907A AGILENT TECHNOLOGIES, 34907A Datasheet - Page 25

Data Acquisition Switch Unit

34907A

Manufacturer Part Number
34907A
Description
Data Acquisition Switch Unit
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 34907A

Input Current Max
400mA
Bandwidth
100kHz
Input Voltage Max
42V
Lead Free Status / RoHS Status
na
34907A
Multifunction Module
• 16 bits of digital input and output
• 100 kHz totalizer input
• Two ± 12 V analog outputs
The Agilent 34907A allows great flexibility for a variety
of sense and control applications. It combines two 8-bit
ports of digital input and output, a 100 kHz gated totalizer,
and two ± 12 V analog outputs—all on a single earth-
referenced module. The digital inputs and totalizer input
may be included in a scan. Alarm limits for the digital and
event counter inputs are evaluated continuously, capturing
and logging alarm conditions even between scans.
26 Bits
8
8
16
16
DAC 1
DAC 2
DIO
TOT
Bit 0
Bit 7
Bit 0
Bit 7
+IN
-IN
Gate
Gate
Channel 04
Channel 05
Port 1
Channel 01
Port 2
Channel 02
Channel 03
Digital input/ouput
Use the digital outputs with an external power supply to
control microwave switches and attenuators, solenoids,
power relays, indicators, and more. Use the digital inputs
to sense limit switch and digital bus status. There are no
complex handshake modes; reads and writes are initiated
either from the front panel or the bus.
Port 1, 2
Vin(L)
Vin(H)
Vout(L)
Vout(H)
Vin(H) max
Alarming
Read/Write Speed 95/s
Totalize input
Count events from devices like photo interrupters, limit
switches, and Hall-effect sensors.
It keeps an updated total which can be read via the front
panel or programmatically at any time. With 26 bits of
resolution, it can count events at full speed for nearly
11 minutes without an overflow.
Max count
Totalize input
Signal level
Threshold
Gate Input
Count reset
Read speed
Analog output
Use the two electronically calibrated analog outputs to
source bias voltages to your device under test, to control
your analog programmable power supplies, or use the
outputs as setpoints for your control systems. The outputs
are programmed directly in volts, either from the front panel
or from the bus.
DAC 1, 2
Resolution
IOUT
Settling time
Accuracy
Temp. coefficient
1 year ± 5 °C
Speed
Latency
8 bit, input or output, nonisolated
< 0.8 V (TTL)
> 2.0 V (TTL)
< 0.8 V @ Iout = -400 mA
> 2.4 V @ Iout = 1 mA
< 42 V with external open drain pull-up
Maskable pattern match or state
change
4 ms (max) alarm sampling
5 ms (typical) to 34970A alarm output
2
100 kHz (max) Rising or falling edge,
programmable
1 Vp-p (min) 42 Vpk (max)
0 V or TTL, jumper selectable
TTL-Hi, TTL-Lo, or none
Manual or Read + Reset
85/s
± 12 V, nonisolated
1 mV
10 mA max
1 ms to 0.01% of output
± (% of output + mV)
0.25% + 20 mV
± ( 0.015% + 1 mV)/°C
26
- 1
25