U3401A-ABA AGILENT TECHNOLOGIES, U3401A-ABA Datasheet - Page 113

no-image

U3401A-ABA

Manufacturer Part Number
U3401A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3401A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3401A Dual Display Multimeter
U3401A User’s and Service Guide
Measurement
Table 6-14 Supplemental measurement specifications
AC current (true
RMS, AC
coupling mode)
AC current (true
RMS, AC+DC
coupling mode)
Resistance
Measurement method
Crest factor
Shunt resistance
Input protection
Reponse time
Measurement method
Crest factor
Measurement range
Response time
Measurement method
Open-circuit voltage
Zeroing error
Input protection
Response time
Specification
DC coupled to the fuse and current shunt, AC
coupled true rms measurement (measures the AC
component only)
Maximum 3:0 at full scale
0.01 Ω to 100 Ω for 500 μA to 10 A ranges
Front panel fuse 630 mA, 500 V, Internal 25 A,
440 V
Approximate 1.5 second when the displayed
reading reaches 99.9% AC value of the tested
input signal at the same range.
AC+DC coupled to the fuse and current shunt,
AC+DC coupled true RMS measurement
(measures the AC component only)
Maximum 3:0 at full scale
VDC and VAC are automatically set at the same
range
Approximate 1.5 second when the displayed
reading reaches 99.9% AC value of the tested
input signal at the same range.
Two-wire Ohms
Limited to +6.0 VDC
0.05 Ω or less(excluding test lead resistances) in
each range when Rel operation is used
500 V DC or AC RMS
Approximately 1.5 seconds for 5 MΩ and ranges
below 5 MΩ.
Specifications and Characteristics