U3401A-ABA AGILENT TECHNOLOGIES, U3401A-ABA Datasheet - Page 82

no-image

U3401A-ABA

Manufacturer Part Number
U3401A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3401A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3401A Dual Display Multimeter
4
64
Performance Test
Resistance Verification Test
Configuration: 2- Wire Ω
1 Select the resistance function.
2 Select each range in the order shown in
Table 4-4 2-wire Ω verification test
Function
2-wire Ω
[1]
resistance.
resistance value indicated. Compare measurement results to the
appropriate test limits shown in
appropriate source settling).
Specifications are for 2-wire ohms function using the Rel operation enabled to eliminate lead
Input
0.00 Ω
0.0000 kΩ
0.000 kΩ
0.00 kΩ
0.0000 MΩ
0.000 MΩ
450.00 Ω
4.5000 kΩ
45.000 kΩ
450.00 kΩ
4.5000 MΩ
30.000 MΩ
Range
500 Ω
5 kΩ
50 kΩ
500 kΩ
5 MΩ
50 MΩ
500 Ω
5 kΩ
50 kΩ
500 kΩ
5 MΩ
50 MΩ
Table 4-
Table 4-
4. (Be certain to allow for
U3401A User’s and Service Guide
Error from nominal one year
±50 mΩ
±0.3 Ω
±3 Ω
±30 Ω
±300 Ω
±3 kΩ
±500 mΩ
±4.8 Ω
±48 Ω
±480 Ω
±4.8 kΩ
±93 kΩ
4. Provide the
[1]
[1]
[1]
[1]

Related parts for U3401A-ABA