U3402A-ABA AGILENT TECHNOLOGIES, U3402A-ABA Datasheet - Page 114

no-image

U3402A-ABA

Manufacturer Part Number
U3402A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3402A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3402A Dual Display Multimeter
6
94
Specifications and Characteristics
DC Current
Table 6-2 DCI resolution, full scale reading, and accuracy [± (% of reading + count)]
Rate
[1]
[2]
Slow
Medium
Fast
Typical at full scale reading and voltage across the input terminals.
Use Rel operation.
Range
12.0000 mA
120.000 mA
1200.00 mA
12.0000 A
40.000 mA
120.00 mA
1200.0 mA
12.000 A
40.00 mA
120.0 mA
1200 mA
12.00 A
Resolution
0.1 µA
1 µA
10 µA
100 µA
1 µA
10 µA
100 µA
1 mA
10 µA
100 µA
1 mA
10 mA
Maximum
reading
11.9999
119.9999
1199.99
11.9999
39.999
119.99
1199.9
11.999
39.99
119.9
1199
11.99
Accuracy
(One year;
23ºC ± 5ºC)
U3402A User’s and Service Guide
0.05% + 15
0.05% + 5
0.2% + 5
0.2% + 5
0.1% + 6
0.1% + 3
0.2% + 3
0.2% + 3
0.1% + 2
0.1% + 2
0.2% + 2
0.2% + 2
[2]
Burden voltage
and shunt resistor
<0.15 V/10
<1.5 V/10
<0.3 V/0.1
<0.6 V/0.01
<0.5 V/10
<1.5 V/10
<0.3 V/0.1
<0.6 V/0.01
<0.5 V/10
<1.5 V/10
<0.3 V/0.1
<0.6 V/0.01
[1]