2
Figure 2-6 2-wire
26
CAU T ION
Operations and Features
Performing Resistance Measurements
terminal connection and display
Disconnect circuit power and discharge all high-voltage capacitors before measuring
resistance to avoid damaging the multimeter or the device under test.
Measuring resistance
1 Press
2 Connect the red and black test leads to the input terminal as shown in
3 Probe the test points (by shunting the resistor) and read the display.
In autoranging mode, the multimeter automatically selects the appropriate
range and the measurement is displayed.
• Seven ranges:
• Measurement method:
• Input protection:
Figure 2- 6
measurement method).
Resistance
or
+
. The default function is 2- wire
Test current
Figure 2- 7
• Slow reading rate: 120 .000 , 1.20000 k , 12.0000 k ,
• Medium reading rate: 400 .00 , 4.0000 k , 40.000 k ,
• Fast reading rate: 400 .0 , 4.000 k , 40.00 k , 400.0 k ,
2-wire ohms or 4-wire ohms, open circuit voltage limited to < 5 V
500 V on all ranges
respectively (according to the selected
120.000 k , 1.20000 M , 12.0000 M , 120.000 M
400.00 k , 4.0000 M , 40.000 M , 300.00 M
4.000 M , 40.00 M , 300.0 M
U3402A User’s and Service Guide
measurement.