Specifications of PAXS0010

No. Of Digits / Alpha
5
Meter Function
DC Millivoltmeter
Meter Range
± 24mV To ± 240mV
Digit Height
14.2mm
Power Consumption
11W
Operating Temperature Range
0°C To +50°C
Supply Voltage Dc, Min
11V
Accuracy
0.07% of Reading +4 µV
Brand/series
PAX
Character Height
0.56 in.
Connection Type
2-Wire (Single-Ended), 4-Wire Bridge (Differential)
Cut Out, Panel
1⁄8 DIN
Dimensions
4.2" L × 3.8" W × 1.95" H
Display Digit Height
0.56 "
Display Resolution
1 μV/10 μV
Display Type
LED
Function
Strain Gage
Input Type
Strain Gauge/Bridge Input
Number Of Digits
5
Primary Type
Electronic
Range, Measurement
±24/±240 mvDC
Special Features
Programmable Function Keys
Temperature Range
0 to +50 °C
Temperature, Storage
–40 to +60 °C
Termination
Cage Clamp
Voltage, Input
11-36 VDV, 24 VAC
Lead Free Status / RoHS Status
na
1.
2. POWER:
3. ANNUNCIATORS:
4. KEYPAD: 3 programmable function keys, 5 keys total
5. A/D CONVERTER: 16 bit resolution
6. UPDATE RATES:
7. DISPLAY MESSAGES:
8. INPUT CAPABILITIES: See specific product specifications, pages 4-6
9. EXCITATION POWER: See specific product specifications, pages 4-6
10. LOW FREQUENCY NOISE REJECTION: (Does not apply to PAXH)
11. USER INPUTS: Three programmable user inputs
12. TOTALIZER:
G
DISPLAY: 5 digit, 0.56" (14.2 mm) red sunlight readable or standard green
LEDs, (-19999 to 99999)
AC Versions:
DC Versions (Not available on PAXH):
MAX - maximum readout selected
MIN - minimum readout selected
TOT - totalizer readout selected, flashes when total overflows
SP1 - setpoint alarm 1 is active
SP2 - setpoint alarm 2 is active
SP3 - setpoint alarm 3 is active
SP4 - setpoint alarm 4 is active
Units Label - optional units label backlight
A/D conversion rate: 20 readings/sec.
Step response: 200 msec. max. to within 99% of final readout value
Display update rate: 1 to 20 updates/sec.
Setpoint output on/off delay time: 0 to 3275 sec.
Analog output update rate: 0 to 10 sec
Max./Min. capture delay time: 0 to 3275 sec.
“OLOL” - Appears when measurement exceeds + signal range.
“ULUL” - Appears when measurement exceeds - signal range
PAXT: “SHrt” - Appears when shorted sensor is detected. (RTD only)
PAXT: “OPEN” - Appears when open sensor is detected.
“. . . .” - Appears when display values exceed + display range.
“- . . .” - Appears when display values exceed - display range.
“E . . .” - Appears when Totalizer exceeds 9 digits.
“h . . .” - Denotes the high order display of the Totalizer.
Normal Mode: > 60 dB @ 50 or 60 Hz ±1%, digital filter off
Common Mode: >100 dB, DC to 120 Hz
Max. Continuous Input: 30 VDC
Isolation To Sensor Input Common: Not isolated. (Not PAXH)
Response Time: 50 msec. max.
Logic State: Jumper selectable for sink/source logic
Function:
Time Accuracy: 0.01% typical
Decimal Point: 0 to 0.0000
Scale Factor: 0.001 to 65.000
Low Signal Cut-out: -19,999 to 99,999
Total: 9 digits, display alternates between high order and low order readouts
AC Power: 85 to 250 VAC, 50/60 Hz, 15 VA
Isolation: 2300 Vrms for 1 min. to all inputs and outputs.
DC Power: 11 to 36 VDC, 11 W
AC Power: 24 VAC, ± 10%, 50/60 Hz, 15 VA
Isolation: 500 Vrms for 1 min. to all inputs and outputs (50 V working).
(digital filter and internal zero correction disabled)
700 msec. max. (digital filter disabled, internal zero correction enabled)
PAXH Only: 1 sec max. to within 99% of final readout value (digital filter
PAXH: Isolation to Sensor Input Common: 1400 Vrms for 1 min.
Time Base: second, minute, hour, or day
Batch: Can accumulate (gate) input display from a user input
INPUT STATE
ENERAL
Inactive
(derate operating temperature to 40° C if operating <15 VDC and three
disabled)
Working Voltage: 125 V
Active
plug-in option cards are installed)
22 KΩ Ω pull-up to +5 V
SINKING INPUTS
V
V
IN
IN
M
< 0.9 VDC
> 3.6 VDC
ETER
SOURCING INPUTS
22 KΩ Ω pull-down
V
V
S
IN
IN
> 3.6 VDC
< 0.9 VDC
PECIFICATIONS
3
13. CUSTOM LINEARIZATION:
14. MEMORY: Nonvolatile E
15. ENVIRONMENTAL CONDITIONS:
16. CERTIFICATIONS AND COMPLIANCES:
17. CONNECTIONS: High compression cage-clamp terminal block
18. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 outdoor use.
19. WEIGHT: 10.4 oz. (295 g)
Data Point Pairs: Selectable from 2 to 16
Display Range: -19,999 to 99,999
Decimal Point: 0 to 0.0000
PAXT: Ice Point Compensation: user value (0.00 to 650.00 μV/°C)
and display values.
Operating Temperature Range: 0 to 50°C (0 to 45°C with all three plug-in
Vibration According to IEC 68-2-6: Operational 5 to 150 Hz, in X, Y, Z
Shock According to IEC 68-2-27: Operational 25 g (10g relay), 11 msec in 3
Storage Temperature Range: -40 to 60°C
Operating and Storage Humidity: 0 to 85% max. RH non-condensing
Altitude: Up to 2000 meters
SAFETY
ELECTROMAGNETIC COMPATIBILITY
Notes:
1. Self-recoverable loss of performance during EMI disturbance at 10 V/m:
Refer to EMC Installation Guidelines section of the bulletin for additional
Wire Strip Length: 0.3" (7.5 mm)
Wire Gage: 30-14 AWG copper wire
Torque: 4.5 inch-lbs (0.51 N-m) max.
IP20 Touch safe. Installation Category II, Pollution Degree 2. One piece
bezel/case. Flame resistant. Synthetic rubber keypad. Panel gasket and
mounting clip included.
Immunity to EN 50082-2
Electrostatic discharge
Electromagnetic RF fields
Fast transients (burst)
RF conducted interference
Simulation of cordless telephones
Emissions to EN 50081-2
RF interference
cards installed)
direction for 1.5 hours, 2g’s.
directions.
UL Recognized Component, File #E179259, UL61010A-1, CSA C22.2
No. 61010-1
UL Listed, File # E137808, UL508, CSA C22.2 No. 14-M95
Type 4X Enclosure rating (Face only), UL50
IECEE CB Scheme Test Certificate #US/8843A/UL
For operation without loss of performance:
information.
PAXT Only: File # E156876, UL873, CSA C22.2 No. 24
Recognized to U.S. and Canadian requirements under the Component
Recognition Program of Underwriters Laboratories, Inc.
LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards
CB Scheme Test Report #04ME11209-20041018
IEC 61010-1, EN 61010-1: Safety requirements for electrical
IP65 Enclosure rating (Face only), IEC 529
IP20 Enclosure rating (Rear of unit), IEC 529
Measurement input and/or analog output signal may deviate during
Unit is mounted in a metal enclosure (Buckeye SM7013-0 or equivalent)
I/O and power cables are routed in metal conduit connected to earth
Issued by Underwriters Laboratories, Inc.
equipment for measurement, control, and laboratory use, Part I
EMI disturbance.
ground.
2
PROM retains all programmable parameters
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-6
ENV 50204
EN 55011
Level 2; 4 Kv contact
Level 3; 8 Kv air
Level 3; 10 V/m
80 MHz - 1 GHz
Level 4; 2 Kv I/O
Level 3; 2 Kv power
Level 3; 10 V/rms
150 KHz - 80 MHz
Level 3; 10 V/m
900 MHz ±5 MHz
200 Hz, 50% duty cycle
Enclosure class A
Power mains class A
1